Popis: |
Dedicated cardiac SPECT scanners have improved performance over standard gamma cameras allowing reductions in acquisition times and/or injected activity. One approach to improving performance has been to use pinhole collimators, but this can cause position-dependent variations in attenuation, sensitivity, and spatial resolution. CT attenuation correction (AC) and an accurate system model can compensate for many of these effects; however, scatter correction (SC) remains an outstanding issue. In addition, in cameras using cadmium-zinc-telluride-based detectors, a large portion of unscattered photons is detected with reduced energy (low-energy tail). Consequently, application of energy-based SC approaches in these cameras leads to a higher increase in noise than with standard cameras due to the subtraction of true counts detected in the low-energy tail. Model-based approaches with parallel-hole collimator systems accurately calculate scatter based on the physics of photon interactions in the patient and camera and generate lower-noise estimates of scatter than energy-based SC. In this study, the accuracy of a model-based SC method was assessed using physical phantom studies on the GE-Discovery NM530c and its performance was compared to a dual energy window (DEW)-SC method.The analytical photon distribution (APD) method was used to calculate the distribution of probabilities that emitted photons will scatter in the surrounding scattering medium and be subsequently detected. APD scatter calculations forAPD-modeled projections showed good agreement with SPECT measurements and had reduced noise compared to DEW scatter estimates. APD-SC reduced mean error in activity measurement compared to DEW-SC in images and the reduction was statistically significant where the scatter fraction (SF) was large (mean SF = 28.5%, T-test p = 0.007). APD-SC reduced measurement uncertainties as well; however, the difference was not found to be statistically significant (F-test p0.5). RMSE comparisons showed that elevated levels of scatter did not significantly contribute to a change in RMSE (p0.2).Model-based APD scatter estimation is feasible for dedicated cardiac SPECT scanners with pinhole collimators. APD-SC images performed better than DEW-SC images and improved the accuracy of activity measurement in high-scatter scenarios. |