Nuclear instrumentsmethods in physics research. Section B, Beam interactions with materials and atoms. 438 |
ISSN: |
0168-583X |
Popis: |
We report on the first several years of operation of our recently installed 250 kV SSAMS at LLNL, purchased to replace our 1-MV AMS system for the measurement of (14)C from labeled biochemical samples. We have modified the ion source region to improve ion output. Additionally, the SSAMS required significant software modifications to the data acquisition system in order to accurately measure (14)C at the high-count rates typically encountered with labeled biochemical samples. We found that the data can be corrected assuming a nonparalyzable dead time response with a single event dead time of 6 µs. Since operation began, we have measured over 13,000 graphitic unknowns and over 1900 standards with an overall precision of 1.0%. We have optimized our system for the analysis of CO2 gas samples. We compared aliquots of identical samples measured as solid graphite and as liquid drops. Excellent agreement was found between the two, although the average precision of the graphite targets was an order of magnitude better than the liquid drop analysis due to the much larger number of (14)C atoms available for measurement. |