Autor: |
Zhanshuo, Wang, Wei, Xu, Zhiwei, Hu, Jiangang, Yang, Shan, Gao |
Rok vydání: |
2018 |
Předmět: |
|
Zdroj: |
Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation. 42(3) |
ISSN: |
1671-7104 |
Popis: |
Distortion will occur when testing medical microwave products, due to its high power, high frequency, in the existing radiated emission testing process which will cause saturation non-linear of the pre-amplifier. It will lead to test distorted, the test results are not accurate. This article will improve the original test method by adding some suppressing-frequency devices to suppress the main frequency energy and avoid pre-amplifier saturation, eliminate distortion. All of this will ensure the test results are effective and relia ble. |
Databáze: |
OpenAIRE |