Autor: |
P S, Unwin, S R, Cannon, R J, Grimer, H B, Kemp, R S, Sneath, P S, Walker |
Rok vydání: |
1996 |
Předmět: |
|
Zdroj: |
The Journal of bone and joint surgery. British volume. 78(1) |
ISSN: |
0301-620X |
Popis: |
We have made a retrospective study of 1001 custom-made prostheses used as replacements after surgery for bone tumours. There were 493 distal femoral, 263 proximal femoral and 245 proximal tibial prostheses. Aseptic loosening was shown to be the principal mode of failure of the implants, and 71 patients had revision for aseptic loosening of a cemented intramedullary stem. The probability of a patient surviving aseptic loosening for 120 months was 93.8% for a proximal femoral replacement, 67.4% for a distal femoral prosthesis and 58% for a proximal tibial implant. In patients with distal femoral replacements the age of the patient at the time of operation and the percentage of bone resected were related to the risk of aseptic loosening. Young patients with distal femoral prostheses in whom a high percentage of the femur had been replaced had the poorest prognosis for survival without aseptic loosening. The percentage of bone removed had a significant effect in the proximal tibial replacement group, but the age of the patient did not. By contrast, neither the age nor the percentage of bone removed was a factor after proximal femoral replacement. The significance of these findings is discussed in relation to mechanical factors. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|