Considerations when measuring urinary albumin: precision, substances that may interfere, and conditions for sample storage

Autor: M L, MacNeil, P W, Mueller, S P, Caudill, K K, Steinberg
Rok vydání: 1991
Předmět:
Zdroj: Clinical chemistry. 37(12)
ISSN: 0009-9147
Popis: The measurement of small but abnormal amounts of albumin in urine is important in evaluating kidney disease in people with diabetes mellitus, hypertension, or possible adverse health effects from exposure to nephrotoxins. Routine laboratory methods for measuring albumin are not sensitive enough to measure the amounts that are significant in urine (less than 30 mg/L). In our laboratory we used three immunoassays for measuring urinary albumin: enzyme-linked immunosorbent assay (EIA), radioimmunoassay (RIA), and immunoturbidimetric assay (IT). We calculated the CVs of the three methods, investigated potential interfering substances at three times their normal concentrations, and stored urine under different conditions to find the best way to protect the sample until assay. The potential interferents we checked were transferrin, urea, beta 2-microglobulin, retinol-binding protein, creatinine, kappa and lambda light chains, IgG, hemoglobin, ketone, and glucose. The stability study involved two study temperatures (-20 and -70 degrees C) and four treatments (centrifuging or filtering, before or after storage). We found the following: the RIA had the lowest CV; the results from the interference study showed no interference from normal physiological concentrations of the substances investigated; storage at -70 degrees C regardless of the treatment should be adequate to prevent loss of albumin immunoreactivity.