Cardiovascular journal of South Africa : official journal for Southern Africa Cardiac Society [and] South African Society of Cardiac Practitioners. 15(2) |
Popis: |
To assess the efficacy and safety of single detachable coils in the closure of PDAs.Review of cases in which a single detachable coil was implanted. The diameter of the ampulla was used to select coil size. Diameters of the PDA, ampulla and residual shunts were measured.There were 36 patients with a mean minimal ductal diameter of 2.2 +/- 0.6 mm. Immediately after implantation, 46% of the patients had residual shunts and after 24 hours, only 28%. A final spontaneous closure rate of 94% was observed after 21 months. There was a significant (p0.01) difference when minimal PDA diameter of those who had complete closure within 24 hours (median: 2 mm, interquartile range: 1.7-2.3 mm) was compared to those with a residual shunt (median: 2.5 mm, interquartile range: 2.3-3.2 mm). Negligible complications were experienced.Single Cook detachable coils are effective for PDA closure with a low complication rate if properly selected. There is a high rate of spontaneous closure of trivial residual shunts. In patients with a minimal ductal diameteror = 2 mm and a residual shunt, more coils may be considered. |