Autor: |
B, Vandendorpe, P, Guilbert, C, Champagne, T, Antoni, T D, Nguyen, N, Gaillot-Petit, S, Servagi Vernat |
Jazyk: |
francouzština |
Rok vydání: |
2017 |
Předmět: |
|
Zdroj: |
Cancer radiotherapie : journal de la Societe francaise de radiotherapie oncologique. 21(8) |
ISSN: |
1769-6658 |
Popis: |
To evaluate the dosimetric contribution of helical tomotherapy for breast cancers compared with conformal radiotherapy in mono-isocentric technique.For 23 patients, the dosimetric results in mono-isocentric 3D conformational radiotherapy did not satisfy the constraints either of target volumes nor organs at risk. A prospective dosimetric comparison between mono-isocentric 3D conformational radiotherapy and helical tomotherapy was therefore carried out.The use of helical tomotherapy showed a benefit in these 23 patients, with either an improvement in the conformity index or homogeneity, but with an increase in low doses. Of the 23 patients, two had pectus excavatum, five had past thoracic irradiation and two required bilateral irradiation. The other 14 patients had a combination of morphology and/or indication of lymph node irradiation. For these patients, helical tomotherapy was therefore preferred to mono-isocentric 3D conformational radiotherapy.Tomotherapy appears to provide better homogeneity and tumour coverage. This technique of irradiation may be justified in the case of morphological situations such as pectus exavatum and in complex clinical situations. In other cases, conformal radiotherapy in mono-isocentric technique remains to be favoured. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|