Evaluation of contrast and denoising effects related to imaging parameters of compressed sensitivity encoding in contrast-enhanced magnetic resonance imaging

Autor: Yasuo, Takatsu, Masafumi, Nakamura, Takanobu, Yamashiro, Atsushi, Ikemoto, Satoshi, Sawa, Masanobu, Nakamura, Tosiaki, Miyati
Rok vydání: 2020
Předmět:
Zdroj: Radiological physics and technology. 14(2)
ISSN: 1865-0341
Popis: To acquire reference data for setting an appropriate compressed sensitivity encoding (CS) for brain lesion detectability, the effects of contrast and noise on contrast-enhanced magnetic resonance imaging (MRI) were evaluated. Gadobutrol at various concentrations and manganese chloride tetrahydrate were used as a phantom. Various CS factors (0-10) and denoising levels (weak, medium, and strong) were assessed. The contrast amount decreased from CS7 in non-denoised images for 0.5-2 mmol/L solutions but slightly decreased from CS7 with denoising. The noise amount significantly increased with an increasing CS factor. Generally, there was a significant difference in the denoising level and rate across all CS factors in the case of the 2 and 0 mmol/L solutions. When the CS factor was increased without denoising, the integrated noise power spectrum (NPS) increased and decreased in the high-frequency and low-frequency areas, respectively. These data can be used to establish settings based on the degree of denoising.
Databáze: OpenAIRE