Analysis of normal lung irradiation in radiosurgery treatments: a comparison of lung optimized treatment (LOT) on cyberknife, 4D target volume on helical tomotherapy, and DIBH on linear accelerator

Autor: Raghavendra, Holla, D, Khanna, V K Sathiya, Narayanan, Deb Narayan, Dutta
Rok vydání: 2020
Předmět:
Zdroj: Physical and engineering sciences in medicine. 44(4)
ISSN: 2662-4737
Popis: Quantitative retrospective analysis of the normal lung irradiation due to the variations of the ITV volume based on the techniques used for upper lobe (UL), mid lobe (ML), and lower lobe (LL) lung tumours when used with 2-view, 1-view, 0-view based LOT technique on Cyberknife, AveIP on Helical Tomotherapy, and DIBH on VMAT systems. In the treatment of lung tumours, patients medically inoperable or those who are unwilling to undergo surgery have the option to be treated using radiation therapy. There are many motion control techniques available for the treatment of the moving target, such as movement encompassment, respiratory gating, breath-hold, motion reduction, and tumour monitoring. ITV generation is dependent on technique and hence the volume of the PTVs will differ based on the technique used. This study aimed to determine the influence of these ITVs on the irradiated normal lung volume for UL, ML, and LL lung tumours for 23 patients. The mean difference in the PTV volumes generated with the 0-view technique was significant with that of 2-view and DIBH techniques (p-value 0.04). The mean difference in the PTV volumes generated by 2-view and DIBH was small for UL, ML, and LL tumours. V
Databáze: OpenAIRE