Simultaneous quadruple-label fluorometric immunoassay of thyroid-stimulating hormone, 17 alpha-hydroxyprogesterone, immunoreactive trypsin, and creatine kinase MM isoenzyme in dried blood spots
Autor: | Y Y, Xu, K, Pettersson, K, Blomberg, I, Hemmilä, H, Mikola, T, Lövgren |
---|---|
Rok vydání: | 1992 |
Předmět: |
Samarium
Adrenal Hyperplasia Congenital Cystic Fibrosis 17-alpha-Hydroxyprogesterone Fluoroimmunoassay Infant Newborn Antibodies Monoclonal Thyrotropin Muscular Dystrophies Isoenzymes Neonatal Screening Europium Hypothyroidism Dysprosium Hydroxyprogesterones Humans Indicators and Reagents Trypsin Terbium Creatine Kinase |
Zdroj: | Clinical chemistry. 38(10) |
ISSN: | 0009-9147 |
Popis: | We describe a quadruple-label fluorometric immunoassay for simultaneously measuring four analytes: thyroid-stimulating hormone (TSH), 17 alpha-hydroxyprogesterone (17 alpha-OHP), immunoreactive trypsin (IRT), and creatine kinase MM (CK-MM). The assay is based on immunoreagents labeled with four different lanthanide ions (Eu3+, Tb3+, Sm3+, and Dy3+), on dissociative fluorescence enhancement applying the principle of co-fluorescence, and on time-resolved fluorometry. The monoclonal anti-alpha-TSH and anti-IRT antibodies and the polyclonal anti-CK-MM antibody were labeled with Eu3+, Sm3+, and Dy3+, respectively; 17 alpha-OHP was labeled with Tb3+. The assay was performed in microtitration strip wells coated with a mixture of monoclonal antibodies against beta-TSH, IRT, and CK-MM and a polyclonal goat anti-rabbit IgG for capture of the rabbit anti-17 alpha-OHP antibodies. After completion of the immunoreactions, the bound fractions of the lanthanides were dissociated into the co-fluorescence enhancement solution, creating highly fluorescent chelates. The four lanthanide-specific signals were subsequently measured in a time-resolved fluorometer. The detection limits of the assay were 0.1 mIU/L for TSH, 2 nmol/L for 17 alpha-OHP, 2 micrograms/L for IRT, and 4 U/L for CK-MM. |
Databáze: | OpenAIRE |
Externí odkaz: |