[Equipment surveillance of implantable catheter ports (1996-1998). Members of Sub-Commissions 4a and 4b for Equipment Surveillance]

Autor: F, Conreux, L, Beydon, D, Safran, J B, Cazalaà
Jazyk: francouzština
Rok vydání: 2000
Předmět:
Zdroj: Annales francaises d'anesthesie et de reanimation. 19(3)
ISSN: 0750-7658
Popis: We analyse all incident reports to the French Health Ministry over the 1996-1998 period, concerning implantable catheter ports. They represent 7% of all reports in the field of anaesthesia and intensive care. Two hundred eleven reports have been analysed. There were 93 catheter ruptures with 77 intravascular migrations. In more than 50% of cases, a pinch-off syndrome mechanism was responsible of these ruptures and a displacement of the O'ring was involved in 25%. Additional minor complications were reported and discussed in regard to literature. We conclude that implantable catheter ports may lead to serious incidents which are preventable in most cases.
Databáze: OpenAIRE