Microsample determination of lovastatin and its hydroxy acid metabolite in mouse and rat plasma by liquid chromatography/ionspray tandem mass spectrometry

Autor: Y, Wu, J, Zhao, J, Henion, W A, Korfmacher, A P, Lapiguera, C C, Lin
Rok vydání: 1997
Předmět:
Zdroj: Journal of mass spectrometry : JMS. 32(4)
ISSN: 1076-5174
Popis: A sensitive and specific method was developed and validated to quantitate lovastatin and its hydroxy acid in mouse and rat plasma. This method employs a solid-phase extraction procedure to isolate lovastatin and its hydroxy acid metabolite from the biological matrices (0.1 ml of mouse or rat plasma). The reconstituted extracts were analyzed by liquid chromatography/ionspray tandem mass spectrometry (LC/MS/MS). Simvastatin and simvastatin hydroxy acid were used as internal standards for lovastatin and lovastatin hydroxy acid, respectively. The assay has a lower limit of quantitation (LLQ) of 0.50 ng ml-1 in mouse and rat plasma for both lovastatin and its hydroxy acid based on 0.1 ml aliquots of plasma. The intra- and inter-assay precision (RSD), calculated from quality control (QC) samples, was7% for lovastatin and6% for lovastatin hydroxy acid in both matrices. The inter-assay accuracy as determined from QC samples was less than 6% for lovastatin and less than 8% for lovastatin hydroxy acid in both matrices. The overall recovery of lovastatin was 54% in mouse plasma and 55% in rat plasma, and the overall recovery of lovastatin hydroxy acid was 100% in mouse plasma and 67% in rat plasma.
Databáze: OpenAIRE