Precise Quantitative Assessment of the Clinical Performances of Two High-Flux Polysulfone Hemodialyzers in Hemodialysis: Validation of a Blood-Based Simple Kinetic Model Versus Direct Dialysis Quantification

Autor: Paik-Seong, Lim, Yuyu, Lin, Minfeng, Chen, Xiaoqi, Xu, Yun, Shi, Sudhir, Bowry, Bernard, Canaud
Rok vydání: 2017
Předmět:
Zdroj: Artificial organs. 42(5)
ISSN: 1525-1594
Popis: Highly permeable dialysis membranes with better design filters have contributed to improved solute removal and dialysis efficacy. However, solute membrane permeability needs to be well controlled to avoid increased loss of albumin that is considered to be detrimental for dialysis patients. A novel high-flux dialyzer type (FX CorDiax; Fresenius Medical Care) incorporating an advanced polysulfone membrane modified with nano-controlled spinning technology to enhance the elimination of a broader spectrum of uremic toxins has been released. The aim of this study was to compare in the clinical setting two dialyzer types having the same surface area, the current (FX dialyzer) and the new dialyzer generation (FX CorDiax), with respect to solute removal capacity over a broad spectrum of markers, including assessment of albumin loss based on a direct dialysis quantification method. We performed a crossover study following an A1-B-A2 design involving 10 patients. Phase A1 was 1 week of thrice-weekly bicarbonate hemodialysis with the FX dialyzer, 4 h per treatment; phase B was performed with a similar treatment regimen but with a new FX CorDiax dialyzer and finally the phase A2 was repeated with FX dialyzer as the former phase. Solute removal markers of interest were assessed from blood samples taken before and after treatment and from total spent dialysate collection (direct dialysis quantification) permitting a mass transfer calculation (mg/session into total spent dialysate/ultrafiltrate). On the blood side, there were no significant differences in the solute percent reduction between FX CorDiax 80 and FX 80. On the dialysate side, no difference was observed regarding eliminated mass of different solutes including β
Databáze: OpenAIRE