[Pediatric fracture diagnosis--ultra-low-dose CT with an effective dose equal to that of radiographs]

Autor: J D, Moritz, B, Hoffmann, D H, Sehr, J, Eggerking, K, Keil, A, Caliebe, J, Dischinger, M, Heller, H, Bolte
Jazyk: němčina
Rok vydání: 2012
Předmět:
Zdroj: RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 184(11)
ISSN: 1438-9010
Popis: Computed tomography (CT) plays an important role in trauma diagnosis in children, especially for complex fractures. The aim of this study was to examine the diagnostic value of ultra-low-dose-CT (ULD-CT) with an effective dose equal to that of radiographs in an experimental study and to compare its results with those of radiographs.Limb bones of dead young pigs served as a model for pediatric bones. A total of 51 fractured and non-fractured bones were examined with a 64 multislice-CT with a standard dose protocol as gold standard, with two ultra-low-dose-protocols, and with standard radiographs with different exposures.In spite of high background noise the examinations of ULD-CT were not adequate only in 2 of 204 cases. ULD-CT was slightly superior to radiographs in detection of fractures. ULD-CT could significantly better characterize the fractures than radiographs. The overall result of ULD-CT was significantly better than that of radiographs with standard exposure.ULD-CT with the effective dose of radiographs is successfully applicable in pediatric fracture diagnosis, and its overall result is significantly better than that of radiographs.
Databáze: OpenAIRE