Autor: |
H, Boulahdour, A, Behar, M J, Haardt, J L, Selam |
Rok vydání: |
1996 |
Předmět: |
|
Zdroj: |
The International journal of artificial organs. 19(12) |
ISSN: |
0391-3988 |
Popis: |
The aim of this study was to develop a diagnostic procedure for pumping unit malfunction by radionuclide imaging (RI) and to validate the method by comparing the results with those obtained using more conventional methods. Fifteen radionuclide investigations were performed in 11 patients with intraperitoneal implantable insulin pumps. One mCi of 99 mTc in 1 ml isotonic sodium chloride was injected into the reservoir. The results based on catheter visualization and peritoneal accumulation were compared blindly to the efficacy of alkaline rinses and laparoscopic findings. In all RI stoppage cases except one alkaline rinses failed to restore flow. Where laparoscopy was performed, comparisons were concordant i.e. no outflow from the tip of the catheter. The RI images obtained were reproduced in vitro using a pump under normal flow conditions and complete proximal and distal catheter obstruction. RI is a safe, quick non invasive method which allows the location of the site of pump/catheter malfunction within a one step procedure and the prediction of the efficacy of sodium hydroxide rinses. |
Databáze: |
OpenAIRE |