Quantification of the impact of MLC modeling and tissue heterogeneities on dynamic IMRT dose calculations
Autor: | I B, Mihaylov, F A, Lerma, M, Fatyga, J V, Siebers |
---|---|
Rok vydání: | 2007 |
Předmět: |
Models
Statistical Radiotherapy Planning Computer-Assisted Reproducibility of Results Radiotherapy Dosage Models Biological Sensitivity and Specificity Organ Specificity Anisotropy Body Burden Scattering Radiation Computer Simulation Radiotherapy Conformal Artifacts Radiometry Monte Carlo Method Algorithms Relative Biological Effectiveness Software |
Zdroj: | Medical physics. 34(4) |
ISSN: | 0094-2405 |
Popis: | This study quantifies the dose prediction errors (DPEs) in dynamic IMRT dose calculations resulting from (a) use of an intensity matrix to estimate the multi-leaf collimator (MLC) modulated photon fluence (DPE(IGfluence) instead of an explicit MLC particle transport, and (b) handling of tissue heterogeneities (DPE(hetero)) by superposition/convolution (SC) and pencil beam (PB) dose calculation algorithms. Monte Carlo (MC) computed doses are used as reference standards. Eighteen head-and-neck dynamic MLC IMRT treatment plans are investigated. DPEs are evaluated via comparing the dose received by 98% of the GTV (GTV D 98%), the CTV D 95%, the nodal D 90%, the cord and the brainstem D 02%, the parotid D 50%, the parotid mean dose (D (Mean)), and generalized equivalent uniform doses (gEUDs) for the above structures. For the MC-generated intensity grids, DPE(IGfluence) is within +/- 2.1% for all targets and critical structures. The SC algorithm DPE(hetero) is within +/- 3% for 98.3% of the indices tallied, and within +/- 3.4% for all of the tallied indices. The PB algorithm DPE(hetero) is within +/- 3% for 92% of the tallied indices. Statistical equivalence tests indicate that PB DPE(hetero) requires a +/- 3.6% interval to state equivalence with the MC standard, while the intervals are1.5% for SC DPE(hetero) and DPE(IGfluence). Overall, these results indicate that SC and MC IMRT dose calculations which use MC-derived intensity matrices for fluence prediction do not introduce significant dose errors compared with full Monte Carlo dose computations; however, PB algorithms may result in clinically significant dose deviations. |
Databáze: | OpenAIRE |
Externí odkaz: |