Autor: |
Normand, Robert, Kristina N, Watt, Sophie, Rochette, Lionel, Desponds, Régis, Vaillant, John A, Rowlands |
Rok vydání: |
2015 |
Předmět: |
|
Zdroj: |
Medical physics. 42(1) |
ISSN: |
2473-4209 |
Popis: |
The authors describe a design for prepatient region of interest attenuators (ROIAs) to reduce dose area product (DAP) for clinical use. The authors describe a model to predict DAP values from x-ray technique parameters recorded during a clinical procedure for image sequences obtained in the presence or absence of ROIAs. The model was developed primarily to determine what the DAP to a patient undergoing cardiac catheterization with a ROIA would have been if no ROIA had been used allowing a determination of DAP reduction.Copper ROIAs with thicknesses that vary gradually so as not to cause significant image artifacts were constructed. X-ray image sequences were acquired on a clinical catheterization system with and without ROIAs with varying x-ray technique parameters. DAP values were measured for all said exposures using an ionization chamber and compared to a model the authors developed.The model can predict DAP values within 3.5% on average with or without ROIAs when compared to ionization chamber measurements.The proposed experimental design is adequate for measuring DAP reductions on the order of 1.5-3.5 that are expected when introducing a ROIA during patient catheterization imaging. |