Electret ion chamber and environmental gamma radiation background: Preliminary results on the optimization of a radon gas measurement system
Autor: | la Verde G., Fronzino T., la Commara M., D'Avino V., Pugliese M. |
---|---|
Přispěvatelé: | la Verde, G., Fronzino, T., la Commara, M., D'Avino, V., Pugliese, M. |
Jazyk: | angličtina |
Rok vydání: | 2020 |
Popis: | A study to evaluate the feasibility of using a system routinely used for measurements of radon gas concentration for measuring the background of gamma radiation was carried out. This approach would reduce the employment of different detectors to a single type, standardizing the methodology adopted and reducing the costs affecting other types of detectors. A 1440 hours long measurement was performed, using an experimental EIC system. The results were compared with measurements provided by a proportional counter and thermoluminescence dosimeters used as a reference. The results suggest the potential of the EIC system for unconventional use for gamma radiation background measurements. |
Databáze: | OpenAIRE |
Externí odkaz: |