Accuracy of torque-limiting devices: A comparative evaluation

Autor: Albayrak H, Gumus HO, Tursun F, Kocaagaoglu HH, Kilinc HI
Jazyk: angličtina
Rok vydání: 2017
Předmět:
Popis: STATEMENT OF PROBLEM: To prevent the loosening of implant screws, clinicians should be aware of the output torque values needed to achieve the desired preload. Accurate torque-control devices are crucial in this regard; however, little information is currently available comparing the accuracy of mechanical with that of electronic torque-control devices. PURPOSE: The purpose of this in vitro study was to identify and compare the accuracy of different types of torque-control devices. MATERIAL AND METHODS: Devices from 5 different dental implant manufacturers were evaluated, including 2 spring-type (Straumann, Implance) mechanical devices (MTLD), 2 friction-type (Biohorizons, Dyna) MTLDs, and 1 (Megagen) electronic torque-control device (ETLD). For each manufacturer, 5 devices were tested 5 times with a digital torque tester, and the average for each device was calculated and recorded. The percentage of absolute deviations from the target torque values (PERDEV) were calculated and compared by using 1-way ANOVA. A 1-sample t test was used to evaluate the ability of each device to achieve its target torque value within a 95% confidence interval for the true population mean of measured values (α=.05 for all statistical analyses). RESULTS: One-way ANOVAs revealed statistically significant differences among torque-control devices (P
Databáze: OpenAIRE