European Journal of Medical Research / In-house 3D-printed surgical guides for osseous lesions of the lower jaw : an experimental study

Autor: Postl, Lukas, Mücke, Thomas, Hunger, Stefan, Bissinger, Oliver, Malek, Michael, Holberg, Svenia, Burgkart, Rainer, Krennmair, Stefan
Jazyk: angličtina
Rok vydání: 2021
Předmět:
DOI: 10.1186/s40001-021-00495-w
Popis: Background: The accuracy of computer-assisted biopsies at the lower jaw was compared to the accuracy of freehand biopsies. Methods: Patients with a bony lesion of the lower jaw with an indication for biopsy were prospectively enrolled. Two customized bone models per patient were produced using a 3D printer. The models of the lower jaw were fitted into a phantom head model to simulate operation room conditions. Biopsies for the study group were taken by means of surgical guides and freehand biopsies were performed for the control group. Results: The deviation of the biopsy axes from the planning was significantly less when using templates. It turned out to be 1.30.6 mm for the biopsies with a surgical guide and 3.91.1 mm for the freehand biopsies. Conclusions: Surgical guides allow significantly higher accuracy of biopsies. The preliminary results are promising, but clinical evaluation is necessary. Version of record