Constant Cardiac Output Monitoring Using the Picco and Lidco Methods Versus Pak in Septic Patients: When to Do Calibration?
Autor: | Mladen Širanović, Josip Kovač, Aleksandar Gopčević, Mijo Kelečić, Marinko Vučić, Nataša Kovač, Bojan Rode |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2011 |
Předmět: | |
Zdroj: | Acta clinica Croatica Volume 50 Issue 2 |
ISSN: | 1333-9451 0353-9466 |
Popis: | The accuracy of cardiac output measurement by two most widely used methods of less invasive hemodynamic monitoring and by the standard technique of thermodilution with pulmonary catheter was assessed. The measurements were carried out in septic surgical patients immediately after and between system calibrations. Study results showed satisfactory compatibility of measurements performed by the two methods and by pulmonary catheter in both phases, thus system calibration being recommendable in hemodynamically unstable septic patients. U ovom radu uspoređivala se točnost mjerenja srčanog izbačaja pomoću dviju najčešće upotrebljavanih metoda manje invazivnog hemodinamskog nadzora i standardnom tehnikom termodilucije plućnim kateterom. Mjerenje je provedeno kod septičnih kirurških bolesnika u razdoblju neposredno nakon i između kalibracija sustava. Rezultati su pokazali zadovoljavajuću podudarnost mjerenja u obje faze primjenom obiju metoda i pomoću plućnog katetera, pa se kalibriranje sustava preporuča u hemodinamski nestabilnih septičnih bolesnika. |
Databáze: | OpenAIRE |
Externí odkaz: |