Autor: |
Poch, F.G.M., Rieder, C., Ballhausen, C., Knappe, V., Ritz, J.-P., Gemeinhardt, O., Kreis, M.E., Lehmann, K.S. |
Přispěvatelé: |
Publica |
Jazyk: |
angličtina |
Rok vydání: |
2016 |
Popis: |
Purpose: Major limitations of conventional RFA are vascular cooling effects. However, vascular cooling effects are supposed to be less pronounced in multipolar RFA. The objective of this ex vivo study was a systematic evaluation of the vascular cooling effects in multipolar RFA. Materials and methods: Multipolar RFA with three bipolar RFA applicators was performed ex vivo in porcine liver (applicator distance 20 mm, energy input 40 kJ). A saline-perfused glass tube ('vessel') was placed parallel to the applicators in order to simulate a natural liver vessel. Five applicator-to-vessel geometries were tested. A liquid-filled glass tube without perfusion was used as a dry run. Ablations were orthogonally cut to the applicators at a defined height. Cooling effects were analysed qualitatively and quantitatively along these cross sectional areas. Results: Thirty-six ablations were performed. A cooling effect could be seen in all ablations with perfused vessels compared to the dry run. While this cooling effect did not have any influence on the ablation areas (859-1072 mm2 versus 958 mm2 in the dry run, p > 0.05), it had a distinctive impact on ablation shape. A vascular cooling effect could be observed in all ablations with perfusion directly around the vessel independent of the applicator position compared to the dry run (p |
Databáze: |
OpenAIRE |
Externí odkaz: |
|