Infrared or ultrasonic milk analysis can affect its results?

Autor: Manske, Guilherme Augusto, Rigo, Elisandra, Gomes, Fábio José, Schogor, Ana Luiza Bachmann
Jazyk: angličtina
Rok vydání: 2019
Předmět:
Zdroj: Ciência Rural, Volume: 49, Issue: 7, Article number: e20180510, Published: 04 JUL 2019
Popis: Constant monitoring of milk quality by the processing industries demands the use of methodologies that add agility and low-cost analysis, such as the use of electronic equipment based on infrared or ultrasonic spectroscopy. In the present study, quality of raw milk received by dairy farms from western region of Santa Catarina state, Brazil, was evaluated throughout two methodologies of analysis of chemical composition and compared, being that both equipment used standard procedures for calibration, considering as reference data from the classic analytical methods recommended in Brazilian legislation. Milk samples from 45 producers were analyzed for SCC, TBC and physicochemical analysis, performed at dairy’s laboratory and at the official laboratory. Infrared and ultrasound methodologies for chemical composition of fat, protein and lactose were compared and correlated. The comparison of the methods showed that there was no significant difference for the mean values of fat (P=0.06); however, they presented significant difference for protein (P=0.001), higher for IR (3.23 vs. 3.33), and lactose (P