Measurement of 14 MeV neutrons at TFTR with Si-diode detectors

Autor: Ruskov, E, Heidbrink, WW, Duong, HH, Roquemore, AL, Strachan, JD
Jazyk: angličtina
Rok vydání: 1995
Předmět:
Zdroj: Ruskov, E; Heidbrink, WW; Duong, HH; Roquemore, AL; & Strachan, JD. (1995). Measurement of 14 MeV neutrons at TFTR with Si-diode detectors. Review of Scientific Instruments, 66(1), 910-912. doi: 10.1063/1.1146202. UC Irvine: Retrieved from: http://www.escholarship.org/uc/item/6dm0b55t
Popis: A detector system based on partially depleted silicon surface barrier detectors and fast front-end electronics has been built and cross calibrated to a set of absolutely calibrated 4He recoil detectors. The cross-calibration factor for the channel with the widest dynamic range is 2.5×10-13 counts per 14 MeV source neutron. These data agree well with the independent neutron activation data. The new detector system covers a large dynamic range (corresponding to 1013-1018 neutrons/s). The response is linear, except at the highest count rates where the detector dead time (∼200 ns) causes departure from linearity. The noise discrimination against 2.5 MeV neutrons and γ pileup is excellent. Measurements of D-T neutrons from a tritium gas puff experiment as well as from a high-power D-T discharge in the TFTR tokamak are presented. © 1995 American Institute of Physics.
Databáze: OpenAIRE