High-power near-concentric Fabry-Perot cavity for phase contrast electron microscopy

Autor: Turnbaugh, Carter, Axelrod, Jeremy J, Campbell, Sara L, Dioquino, Jeske Y, Petrov, Petar N, Remis, Jonathan, Schwartz, Osip, Yu, Zanlin, Cheng, Yifan, Glaeser, Robert M, Mueller, Holger
Rok vydání: 2021
Předmět:
Zdroj: The Review of scientific instruments, vol 92, iss 5
Popis: Transmission electron microscopy (TEM) of vitrified biological macromolecules (cryo-EM) is limited by the weak phase contrast signal that is available from such samples. Using a phase plate would thus substantially improve the signal-to-noise ratio. We have previously demonstrated the use of a high-power Fabry-Perot cavity as a phase plate for TEM. We now report improvements to our laser cavity that allow us to achieve record continuous wave intensities of over 450 GW/cm2, sufficient to produce the optimal 90° phase shift for 300 keV electrons. In addition, we have performed the first cryo-EM reconstruction using a laser phase plate, demonstrating that the stability of this laser phase plate is sufficient for use during standard cryo-EM data collection.