Autor: |
Gu, Y. John, van Oeveren, Willem, Mungroop, Hubert E., Epema, Anne H., den Hamer, Inez J., Keizer, Jorrit J., Leuvenink, Ron P., Mariani, Massimo A., Rakhorst, Gerhard |
Přispěvatelé: |
Cardiovascular Centre (CVC) |
Jazyk: |
angličtina |
Rok vydání: |
2011 |
Předmět: |
|
Zdroj: |
ARTIFICIAL ORGANS, 35(2), E18-E26. Wiley |
ISSN: |
0160-564X |
Popis: |
Although the centrifugal pump has been widely used as a nonpulsatile pump for cardiopulmonary bypass (CPB), little is known about its performance as a pulsatile pump for CPB, especially on its efficacy in producing hemodynamic energy and its clinical effectiveness. We performed a study to evaluate whether the Rotaflow centrifugal pump produces effective pulsatile flow during CPB and whether the pulsatile flow in this setting is clinically effective in adult patients undergoing cardiac surgery. Thirty-two patients undergoing CPB for elective coronary artery bypass grafting were randomly allocated to a pulsatile perfusion group (n = 16) or a nonpulsatile perfusion group (n = 16). All patients were perfused with the Rotaflow centrifugal pump. In the pulsatile group, the centrifugal pump was adjusted to the pulsatile mode (60 cycles/min) during aortic cross-clamping, whereas in the nonpulsatile group, the pump was kept in its nonpulsatile mode during the same period of time. Compared with the nonpulsatile group, the pulsatile group had a higher pulse pressure (P |
Databáze: |
OpenAIRE |