Department of Physics, Iwate Medical University, Department of Nutritional Science, Faculty of Applied Bio-science, Tokyo University of Agriculture, Department of Cardiac Physiology, National Cardiovascular Center Research Institute, Electron Tube Division #2, Hamamatsu Photonics K. K., Department of Radiological Technology, School of Health Sciences, Hirosaki University, Department of Microbiology, School of Medicine, Iwate Medical University, Shock Wave Research Center, Institute of Fluid Science, Tohoku University, Department of Applied Physics and Informatics, Faculty of Engineering. Tohoku Gakuin University |
Jazyk: |
angličtina |
Rok vydání: |
2004 |
Předmět: |
|
Zdroj: |
岩手医科大学教養部研究年報. 39:11-17 |
Popis: |
Energy-selective enhanced K-edge angiography utilizing a conventional x-ray generator is described. The x-ray generator is SOFRON NST-1005, and the maximum tube voltage and current are 100 kV and 5 mA, respectively. In the present research, the tube voltage ranged from 45 to 65kV, and the tube current was regulated to optimum values. The exposure time is controlled in order to obtain optimum x-ray intensity. At a charging voltage of 60 kV, the x-ray intensity rate obtained using an aluminum and a barium sulfate filters were 58.4 and 51.6 μGy/s at 0.7m per pulse, respectively, and the dimensions of the focal spot were approximately 1×1 mm. Angiography was performed using both the aluminum and the barium sulfate filters with a charging voltage of 60 kV. |