128-multidetector CT: For assessment of optimal depth of electrode array insertion in cochlear implant operations
Autor: | Amina Ahmed Sultan, Eman Muhammad Abdelslam, Aya Mohamad Abdelaziz, Mohamed Rashad Ghonim |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2018 |
Předmět: |
lcsh:Medical physics. Medical radiology. Nuclear medicine
medicine.medical_specialty business.industry medicine.medical_treatment lcsh:R895-920 Multidetector ct Implant device 03 medical and health sciences 0302 clinical medicine Cochlear implant Electrode array medicine Radiology Nuclear Medicine and imaging Radiology 030223 otorhinolaryngology Nuclear medicine business 030217 neurology & neurosurgery Statistical correlation |
Zdroj: | The Egyptian Journal of Radiology and Nuclear Medicine, Vol 49, Iss 3, Pp 652-657 (2018) |
Popis: | Objective: To assess the diagnostic reliability of MDCT in pre-operative evaluation of cochlear implant candidates and post-operative, estimation of depth of insertion. Material and methods: The study includes 40 patients (18 males and 22, females); classified into 2 groups: group A (20 patients): cochlear, implant device was Nucleus-22. Group B (20 patients): device was MED-EL.Cochlear length (CL) and cochlear height (CH) were measured pre-operatively by 128-multidetector CT. Electrode length (EL) and insertion, depth angle (α) were measured post-operatively by MDCT. Results: Group A: mean CL was 9.1 mm ± 0.4 SD; mean CH was 4.1 ± 0.3 SD; mean EL was 18 ± 2.7 SD; mean α angle was 299.05 ± 37 SD. Significant, statistical correlation (P |