Analysis of cerebrovascular autoregulation reactivity index electronic monitoring methods

Autor: Ragauskas, A., Kalvaitis, E., Petkus, V., Daubaris, G., Depreitere, B., Kauno technologijos universitetas
Jazyk: angličtina
Rok vydání: 2011
Předmět:
DOI: 10.5755/j01.eee.114.8.687
Popis: Pressure reactivity index (PRx), which is a moving linear correlation (or Pearson’s correlation) between slow waves of ABP and ICP is used as an estimator of the human brain cerebrovascular autoregulation state. Slow ICP and IBP waves were modeled and PRx was calculated by widely used calculation method. It was shown, that existing PRx calculation method gives non- linear PRx distortions comparing with phase shift of ABP and ICP slow waves. It was proposed PRx calculation method, which eliminates non-linear phase demodulator (Pearson’s correlation coefficient calculator), brought in monitoring of human brain cerebrovascular autoregulation.
Slėgio reaktyvumo indeksas (PRx), kuris apskaičiuojamas kaip lėtųjų ABP ir ICP bangų tiesinė koreliacija (arba Pirsono koreliacija), naudojamas žmogaus smegenų kraujotakos autoreguliacijos būsenų dinamikos stebėsenai. Lėtosios ICP ir ABP bangos buvo sumodeliuotos ir PRx buvo apskaičiuotas plačiai taikomais skaičiavimo metodais. Straipsnyje parodyta, kad plačiai taikomas PRx apskaičiavimo metodas nėra tikslus dėl netiesinės apskaičiuotų PRx verčių priklausomybės nuo ABP ir ICP lėtųjų bangų fazių skirtumo. Pasiūlytas naujas, tikslesnis, PRx apskaičiavimo metodas, kuris pašalina netiesinio fazinio demoduliatoriaus (Pirsono koreliacijos koeficiento skaičiuoklio) sukeliamus netiesinius fiziologinio žmogaus smegenų kraujotakos autoreguliacijos matavimo iškraipymus.
Databáze: OpenAIRE