Cerclage wiring technique after proximal femoral fracture in total hip arthroplasty

Autor: Seung-Moo Han, Zair Fishkin, Israel Ziv
Rok vydání: 1999
Předmět:
Zdroj: The Journal of Arthroplasty. 14:98-101
ISSN: 0883-5403
DOI: 10.1016/s0883-5403(99)90209-7
Popis: Cerclage wires have been used to stabilize proximal femoral cracks after stem insertion in cementless total hip arthroplasty. The objective of this study was to determine the optimal number and orientation of cerclage wires necessary to prevent stem subsidence and crack propagation. The crack was stabilized by 1, 2, or 3 wires placed either normal to the femoral neck axis or normal to the crack. The femora were compressed to 2,670 N while measuring crack opening and stem subsidence. Wires placed normal to the crack allowed less stem subsidence by 3.17 mm and less crack opening by 1.55 mm compared with wires placed normal to the neck. The addition of multiple wires reduced subsidence by 50% and reduced crack opening to