Autor: |
Felix Bärenfänger, Jörg Walbersloh, Rachid El Mouden, Fabian Goerg, Andreas Block, Stefan Rohde |
Rok vydání: |
2021 |
Předmět: |
|
Zdroj: |
European journal of radiology. 147 |
ISSN: |
1872-7727 |
Popis: |
A novel two-part protective system consisting of a modified thyroid collar and a head protection is intended to reduce the radiation dose to the examiners head during fluoroscopy-guided interventions. In this pilot study, we tested this protection system under real-life conditions in general radiological and neuroradiological interventions.Two sets of the protection system (set A and B) were equipped with 12 thermoluminiscence detectors (TLD). For simultaneous measurement of radiation exposure and dose-reduction, each six TLDs were fixed to the inner side and on the corresponding outer side of the protection system. Set A was used exclusively for general radiological interventions and set B exclusively for neuroradiological interventions. To compare the staff exposure in general radiology and neuroradiology, dose values were normalized to a DAP of 10 000 µGy∙mThe sets were tested during 20 general radiological interventions and 32 neuroradiological interventions. In neuroradiology, the mean normalized radiation exposure was 13.44 ± 1.36 µSv/10000 µGy∙mFluoroscopy-guided interventions lead to significant radiation exposure of the head area for the examiner. The novel protection system tested led to a significant dose reduction of 80-90%, depending on the type of intervention. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|