Distribution of inhaled nitric oxide during sequential and continuous administration into the inspiratory limb of the ventilator
Autor: | Qin Lu, L. Puybasset, P. Coriat, Jean-Jacques Rouby, G.S. Umamaheswara Rao, Lluis Gallart, E. Mourgeon, J.D. Law-Koune |
---|---|
Rok vydání: | 1997 |
Předmět: |
Artificial ventilation
Mechanical ventilation Analysis of Variance Respiratory Distress Syndrome ARDS Lung Inhalation business.industry medicine.medical_treatment Pressure support ventilation In Vitro Techniques Lung injury Nitric Oxide Critical Care and Intensive Care Medicine medicine.disease Respiration Artificial medicine.anatomical_structure Anesthesia Respiratory Mechanics Humans Medicine business Biotransformation Tidal volume |
Zdroj: | Intensive Care Medicine. 23:849-858 |
ISSN: | 1432-1238 0342-4642 |
DOI: | 10.1007/s001340050421 |
Popis: | The concentrations of nitric oxide (NO) in the ventilatory circuits and the patient's airways were compared between sequential (SQA) and continuous (CTA) administration during inspiratory limb delivery. Design: Prospective controlled study. Setting: 14-bed Surgical Intensive Care Unit of a teaching University hospital. Patients and participants: Eleven patients with acute lung injury on mechanical ventilation and two healthy volunteers. Interventions: A prototype NO delivery device (Opti-NO) and Cesar ventilator were set up in order to deliver 1, 3 and 6 parts per million (ppm) of NO into the bellows of a lung model in SQA and CTA. Using identical ventilatory and Opti-NO settings, NO was administered to the patients with acute lung injury. Measurements and results: NO concentrations measured from the inspiratory limb [INSP-NOMeas] and the trachea [TRACH-NOMeas] using fast response chemiluminescence were compared between the lung model and the patients using controlled mechanical ventilation with a constant inspiratory flow. INSP-NOMeas were stable during SQA and fluctuated widely during CTA (fluctuation at 6 ppm = 61 % in the lung model and 58 ± 3 % in patients). In patients, [TRACH-NOMeas] fluctuated widely during both modes (fluctuation at 6 ppm = 55 ± 3 % during SQA and 54 ± 5 % during CTA). The NO flow requirement was significantly lower during SQA than during CTA (74 ± 0.5 vs 158 ± 2.2 ml.min–1 to attain 6 ppm, p = 0.0001). INSP-NOMeas were close to the values predicted using a classical formula only during SQA (bias = –0.1 ppm, precision = ± 1 ppm during SQA; bias = 2.93 ppm and precision = ± 3.54 ppm during CTA). During SQA, INSP-NOMeas varied widely in healthy volunteers on pressure support ventilation. Conclusions: CTA did not provide homogenous mixing of NO with the tidal volume and resulted in fluctuating INSP-NOMeas. In contrast, SQA delivered stable and predictable NO concentrations during controlled mechanical ventilation with a constant inspiratory flow and was economical compared to CTA. However, SQA did not provide stable and predictable NO concentrations during pressure support ventilation. |
Databáze: | OpenAIRE |
Externí odkaz: |