Usability evaluation of a locomotor therapy device considering different strategies
Autor: | Jörg Schröttner, Christian Baumgartner, Alexander Lassnig, Sonja Langthaler |
---|---|
Rok vydání: | 2016 |
Předmět: |
medicine.medical_specialty
business.industry 010401 analytical chemistry Biomedical Engineering Usability 01 natural sciences usability testing 030218 nuclear medicine & medical imaging 0104 chemical sciences use error 03 medical and health sciences 0302 clinical medicine Physical medicine and rehabilitation operating safety promediks medicine Medicine business Psychology useprob |
Zdroj: | Current Directions in Biomedical Engineering, Vol 2, Iss 1, Pp 67-69 (2016) |
ISSN: | 2364-5504 |
DOI: | 10.1515/cdbme-2016-0018 |
Popis: | Usability of medical devices is one of the main determining factors in preventing use errors in treatment and strongly correlates to patient safety and quality of treatment. This thesis demonstrates the usability testing and evaluation of a prototype for locomotor therapy of infants. Therefore, based on the normative requirements of the EN 62366, a concept combined of evaluation procedures and assessing methods was created to enable extensive testing and analysis of the different aspects of usability. On the basis of gathered information weak points were identified and appropriate measures were presented to increase the usability and operating safety of the locomotor prototype. The overall outcome showed an usability value of 77.4% and an evaluation score of 6.99, which can be interpreted as “satisfactory”. |
Databáze: | OpenAIRE |