Characterization of a multi-pinhole molecular breast tomosynthesis scanner
Autor: | J. Huizenga, Freek J. Beekman, Beien Wang, Jarno van Roosmalen, Marlies C Goorden, R. Kreuger |
---|---|
Rok vydání: | 2020 |
Předmět: |
Scanner
Pinhole collimator Breast imaging Imaging phantom 030218 nuclear medicine & medical imaging law.invention 03 medical and health sciences 0302 clinical medicine Optics law Humans Gamma Cameras Radiology Nuclear Medicine and imaging Breast Radionuclide Imaging Gamma camera Image resolution Tomography Emission-Computed Single-Photon Physics Radiological and Ultrasound Technology Phantoms Imaging business.industry Detector Collimator SPECT 030220 oncology & carcinogenesis Female Pinhole (optics) Radiopharmaceuticals business |
Zdroj: | Physics in Medicine and Biology, 65(19) |
ISSN: | 1361-6560 0031-9155 |
Popis: | In recent years, breast imaging using radiolabelled molecules has attracted significant interest. Our group has proposed a multi-pinhole molecular breast tomosynthesis (MP-MBT) scanner to obtain 3D functional molecular breast images at high resolutions. After conducting extensive optimisation studies using simulations, we here present a first prototype of MP-MBT and evaluate its performance using physical phantoms. The MP-MBT design is based on two opposing gamma cameras that can image a lightly compressed pendant breast. Each gamma camera consists of a 250 × 150 mm2 detector equipped with a collimator with multiple pinholes focusing on a line. The NaI(Tl) gamma detector is a customised design with 3.5 mm intrinsic spatial resolution and high spatial linearity near the edges due to a novel light-guide geometry and the use of square PMTs. A volume-of-interest is scanned by translating the collimator and gamma detector together in a sequence that optimises count yield from the scan region. Derenzo phantom images showed that the system can reach 3.5 mm resolution for a clinically realistic 99mTc activity concentration in an 11-minute scan, while in breast phantoms the smallest spheres visible were 6 mm in diameter for the same scan time. To conclude, the experimental results of the novel MP-MBT scanner showed that the setup had sub-centimetre breast tumour detection capability which might facilitate 3D molecular breast cancer imaging in the future. |
Databáze: | OpenAIRE |
Externí odkaz: |