Improved liquid-liquid extraction with inter-well volume replacement dilution workflow and its application to quantify BMS-927711 in rat dried blood spots by UHPLC-MS/MS
Autor: | Anne-Françoise Aubry, Shenita Basdeo, Qin C Ji, Jianing Zeng, Qianping Peng, Naiyu Zheng, Aida Angeles, Adela Buzescu, Laura Patrone, Mark E. Arnold, Kevin J Trouba |
---|---|
Rok vydání: | 2013 |
Předmět: |
Analyte
Pyridines Clinical Biochemistry Liquid-Liquid Extraction Analytical chemistry Pharmaceutical Science Indicator Dilution Techniques High-performance liquid chromatography Analytical Chemistry Workflow Piperidines Liquid–liquid extraction Limit of Detection Tandem Mass Spectrometry Drug Discovery Animals Spectroscopy Chromatography High Pressure Liquid Chromatography Elution Chemistry Extraction (chemistry) Selected reaction monitoring Reproducibility of Results Dilution Rats Standard curve Dried Blood Spot Testing |
Zdroj: | Journal of pharmaceutical and biomedical analysis. 89 |
ISSN: | 1873-264X |
Popis: | An UHPLC-MS/MS method was developed and validated to quantify BMS-927711, a drug candidate to treat migraine, in rat dried blood spots (DBS). The DBS samples were extracted using an improved liquid-liquid extraction (LLE) strategy involving in the sonication of DBS punches in 20% MeOH aqueous solution containing the internal standard, [(13)C2, D4]-BMS-927711, and then with a 100mM NH4OAc buffer solution, followed by an automated LLE with EtOAc-hexane (70:30, v/v). The presence of 20% MeOH as an organic modifier in the elution solution significantly improved the analyte elution efficiency and assay performance. A novel inter-well volume replacement dilution workflow was introduced for DBS sample dilution before LLE step. This was a simple two-step process, firstly a small portion of the DBS blank solution was discarded, and then the same volume of a concentrated DBS sample solution was spiked into the leftover blank solution to achieve a desired dilution. Chromatographic separation was achieved on an Acuity UPLC(®) BEH C18 column (2.1mm×50mm, 1.7μm) and the analyte was detected by selected reaction monitoring (SRM) with positive electrospray ionization on an AB Sciex Triple Quad 5500 mass spectrometer. The standard curve was linear from 5.00 to 5000ng/mL with assay precision ≤4.9% CV, and assay accuracy within ±3.1%Dev of the nominal values. Accurate sample dilution was achieved by using inter-well volume replacement with a precision of ≤4.2% CV and an accuracy of ±3.3% for dilution QC at 50,000ng/mL with 100-fold dilution (n=18). This robust UHPLC-MS/MS assay has been successfully applied to the non-clinical studies in rats. By using inter-well volume replacement workflow, accurate dilution was demonstrated using only one DBS blank sample for a typical dilution of |
Databáze: | OpenAIRE |
Externí odkaz: |