The repeatability of T2 relaxation time measurement of human knee articular cartilage
Autor: | Eveliina Lammentausta, Osmo Tervonen, Miika T. Nieminen, Ilkka Hannila |
---|---|
Rok vydání: | 2015 |
Předmět: |
Adult
Cartilage Articular Male Reproducibility Materials science Knee Joint Radiological and Ultrasound Technology Coefficient of variation Cartilage Biophysics Reproducibility of Results Delayed Gadolinium Enhanced Magnetic Resonance Imaging of Cartilage Repeatability Magnetic Resonance Imaging Sagittal plane medicine.anatomical_structure Region of interest T2 relaxation Image Processing Computer-Assisted medicine Humans Female Radiology Nuclear Medicine and imaging Biomedical engineering |
Zdroj: | Magnetic Resonance Materials in Physics, Biology and Medicine. 28:547-553 |
ISSN: | 1352-8661 0968-5243 |
DOI: | 10.1007/s10334-015-0494-3 |
Popis: | To assess short- and long-term repeatability of T2 relaxation time measurements of the knee articular cartilage. The right knees of nine asymptomatic volunteers (age 30–38 years, five male, four female) were imaged at 1.5 T in three sessions 1 and 2 weeks apart. To observe short-term repeatability, the measurements were repeated three times within one of the three imaging sessions for each volunteer. T2 relaxation time was mapped using a multi-slice multi-echo spin echo sequence in axial and sagittal planes. Cartilage was manually segmented and repeatability, as measured by root-mean-square coefficient of variation (CVRMS) was evaluated both for the entire bulk cartilage of each joint surface in the slice and separately for each region of interest (ROI) at different topographical locations and separately for the superficial and deep half of each ROI. For bulk T2, the long-term repeatability was 3.2, 5.4, and 3.7 %, and the short-term reproducibility was 3.9, 3.9, and 3.4 % for bulk femoral, tibial, and patellar cartilage, respectively. There were no significant differences between long-term and short-term repeatability in superficial or deep cartilage when comparing CVRMS values (p = 0.338 and 0.700, respectively). For individual ROIs, the repeatability varied between 2.5 and 22.2 % depending on the topographical location. The current results show mostly good repeatability. However, there were remarkable variations of T2 between bulk cartilage and different ROIs, bulk cartilage showing better repeatability. With careful patient positioning T2 can be accurately determined for different cartilage surfaces. |
Databáze: | OpenAIRE |
Externí odkaz: |