First Experience DaTSCAN Imaging Using Cadmium-Zinc-Telluride Gamma Camera SPECT
Autor: | Bernard Songy, Xavier Poullias, Karim Farid, David Lussato, Slavomir Petras, Nadine Caillat-Vigneron, Mohamed Guernou, Mathieu Queneau |
---|---|
Rok vydání: | 2012 |
Předmět: |
Male
Image quality law.invention chemistry.chemical_compound law Spect imaging Humans Medicine Gamma Cameras Radiology Nuclear Medicine and imaging Aged Gamma camera Tomography Emission-Computed Single-Photon Essential tremor business.industry General Medicine Middle Aged medicine.disease Cadmium zinc telluride Zinc chemistry Acquisition time Tellurium business Nuclear medicine Cadmium Tropanes |
Zdroj: | Clinical Nuclear Medicine. 37:e211-e212 |
ISSN: | 0363-9762 |
DOI: | 10.1097/rlu.0b013e3182443f47 |
Popis: | We report our first experience of brain DaTSCAN SPECT imaging using cadmium-zinc-telluride gamma camera (CZT-GC) in 2 cases: a 64-year-old patient suffering from essential tremor and a 73-year-old patient presenting with atypical bilateral extrapyramidal syndrome. In both cases, 2 different acquisitions were performed and compared, using a double-head Anger-GC, followed immediately by a second acquisition on CZT-GC. There were no significant visual differences between images generated by different GC. Our first result suggests that DaTSCAN SPECT is feasible on CZT-GC, allowing both injected dose and acquisition time reductions without compromising image quality. This experience needs to be evaluated in larger series. |