Determining the stability of complete blood count parameters in stored blood samples using the SYSMEX XE-5000 automated haematology analyser

Autor: Wendy McVicker, Emmanuel J. Favaloro, Thirumahal Vanniasinkam, A. Joshi, R. Segalla, V. Luu
Rok vydání: 2015
Předmět:
Zdroj: International Journal of Laboratory Hematology. 37:705-714
ISSN: 1751-5521
Popis: SummaryIntroduction In this study, changes that occur in various blood parameters as determined by the Sysmex XE-5000 analyser upon storage of blood samples for 72 h were investigated. Methods Blood specimens (200) were processed through the SYSMEX XE-5000 haematology analyser within 4 h of collection. Each specimen was distributed into two aliquots and one stored at 4 °C and the other at room temperature (25 ± 1 °C). Both stored aliquots were retested after 24, 48 and 72 h. The mean, mean per cent change and mean absolute difference between the value at 4 h and each time point for all parameters were calculated. Results Among CBC parameters tested, the white cell count, red cell count and haemoglobin levels were found to be stable for up to 72 h. The mean cell volume and haematocrit changed significantly following 24-h storage at room temperature. Reticulocytes were stable for 72 h under both storage conditions. Among the differential parameters, results of the monocyte count displayed significant change after 24-h storage at room temperature. Conclusion The data presented suggest that clinically reliable results for some CBC parameters can be obtained from specimens that are stored at 4 °C for up to 72 h.