Finite element analysis of Stryker Xia pedicle screw in artificial bone samples with and without supplemental cement augmentation
Autor: | Theodore J. Choma, Ferris M. Pfeiffer, Rebecca A. Kueny |
---|---|
Rok vydání: | 2014 |
Předmět: |
Cement
Artificial bone Materials science Lumbar Vertebrae Finite Element Analysis Biomedical Engineering Bone Cements Bioengineering General Medicine Biomechanical testing Prosthesis Design Finite element method Spine Computer Science Applications Biomechanical Phenomena Human-Computer Interaction Pedicle Screws Osteoporotic bone Humans Osteoporosis Cement augmentation Pedicle screw Instant centre of rotation Biomedical engineering |
Zdroj: | Computer methods in biomechanics and biomedical engineering. 18(13) |
ISSN: | 1476-8259 |
Popis: | A validated, using in vitro biomechanical testing, finite element model was used to evaluate the affects of (1) cement augmentation and (2) an intact posterior cortex in osteoporotic bone. The presence of augmentation and/or a posterior cortical cortex increased the stabilization of the pedicle screw 2–5 fold. Placement of cement influenced failure load and toggle; with distal placement having the largest increase in failure load and decrease in cephalad–caudad toggle. The presence of posterior cortex caused a decrease in the amount of toggle, a proximal shift of the center of rotation and an increase in the maximum failure force. |
Databáze: | OpenAIRE |