Objective To assess the biometric features of keratoconic eyes using the Lenstar LS900 and Pentacam systems relative to healthy myopic eyes. Materials and Methods Seventy-three eyes of keratoconic subjects and 83 eyes of control subjects were enrolled. To evaluate the reproducibility of the Lenstar and Pentacam devices’ measurements, keratometric readings [in flattest meridian (Kf), in steepest meridian (Ks), and mean (Km)], central corneal thickness (CCT), and anterior chamber depth (ACD) were obtained using both systems. Axial length and lens thickness (LT) were measured by the Lenstar. The compatibility between the two devices was investigated using the Bland-Altman statistical method. Results Axial length was longer in the myopic group than in eyes with keratoconus (24.94 ± 0.7 and 23.88 ± 0.96 mm, respectively, p Conclusions Axial length was longer in the myopic eyes due to the differences starting from the lens and extending to the posterior segment. Lenstar and Pentacam can be used interchangeably for Km, Kf, and ACD in the myopic group and only for ACD in the keratoconus group. |