Analysis of Low-Level Laser Radiation Transmission in Occlusive Dressings

Autor: Elaine Caldeira de Oliveira Guirro, Fabiana Roberta Nunes, Carla Campos Martins, Rinaldo Roberto de Jesus Guirro
Rok vydání: 2010
Předmět:
Zdroj: Photomedicine and Laser Surgery. 28:459-463
ISSN: 1557-8550
1549-5418
DOI: 10.1089/pho.2009.2524
Popis: The purpose of this study is to analyze the power transmitted by low-level laser therapy (LLLT) into occlusive dressings using different wavelengths for the treatment of cutaneous lesions.LLLT has been largely used to treat several cutaneous lesions commonly associated with occlusive dressings to accelerate the healing process.Radiation transmission was measured by a digital power analyzer connected to a laser emitter with wavelengths of 660, 830, and 904 nm and mean levels of 30, 30, 6.5 mW, respectively, previously calculated. Thirteen different occlusive dressings were analyzed and interposed between the laser emitter and the power analyzer sensor, with 15 measurements made for each dressing. Statistics were provided by the analysis of variance (ANOVA), followed by Student's t-test (p0.05).The power transmitted ranged between 98.6% and 0%, depending on the material and wavelength. The dressings tested were BioFill, Hydrofilm, Confeel Plus 3533, Confeel 3218, DuoDERM Extra Thin, Hydrocoll, Micropore Nexcare, CIEX tape, Emplasto Sábia, CombiDERM, Band-aid, Actisorb Plus, in addition to polyvinylchloride (PVC) film, and transmitted power higher than 40% of the incident power, independently from the wavelength indicated for the association with LLLT.The results showed that LLLT transmission depends on the occlusive dressing material and the wavelength irradiated.
Databáze: OpenAIRE