A fully ingrowing implant for cranial reconstruction: Results in critical size defects in sheep using 3D-printed titanium scaffold

Autor: U. Hubbe, S. Beiser, S. Kuhn, T. Stark, A. Hoess, H. Cristina-Schmitz, I. Vasilikos, M.C. Metzger, R. Rothweiler
Rok vydání: 2021
Předmět:
Zdroj: Biomaterials advances. 136
ISSN: 2772-9508
Popis: Current alloplastic materials such as PMMA, titanium or PEEK don't show relevant bone ingrowth into the implant when used for cranioplasty, ceramic implants have the drawback being brittle. New materials and implant designs are urgently needed being biocompatible, stable enough for cranioplasty and stimulating bone formation. In an in vivo critical size sheep model circular cranial defects (2.4 cm) were covered with three different types of a 3D-printed porous titanium scaffolds with multidirectional, stochastically distributed architecture (uncoated scaffold, hydroxyapatite-coated scaffold, uncoated scaffold filled with a calcium phosphate bone cement paste containing β-TCP granules). An empty titanium mesh served as control. Among the different investigated setups the hydroxyapatite-coated scaffolds showed a surprisingly favourable performance. Push-out tests revealed a 2.9 fold higher force needed in the hydroxyapatite-coated scaffolds compared to the mesh group. Mean CT density at five different points inside the scaffold was 2385HU in the hydroxyapatite-coated group compared to 1978HU in the uncoated scaffold at nine months. Average lateral bone ingrowth after four months in the hydroxyapatite-coated scaffold group was up to the implant center, 12.1 mm on average, compared to 2.8 mm in the control group covered with mesh only. These properties make the investigated scaffold with multidirectional, stochastically distributed structure superior to all products currently on the market. The study gives a good idea of what future materials for cranioplasty might look like.
Databáze: OpenAIRE