Unrecognized hazards of surgical electrodes passed through metal suction-irrigation devices
Autor: | C. R. Voyles, Robert D. Tucker |
---|---|
Rok vydání: | 1994 |
Předmět: |
Suction (medicine)
medicine.medical_specialty Electrosurgery business.industry medicine.medical_treatment Electric Conductivity Suction equipment and supplies Monopolar electrosurgery Cannula Laparoscopes Surgery Models Structural surgical procedures operative Metals Electrode Medicine Laparoscopy business Burns Therapeutic Irrigation Electrodes |
Zdroj: | Surgical endoscopy. 8(3) |
ISSN: | 0930-2794 |
Popis: | Surgical electrodes, passed through metal suction-irrigation devices, pose significant risks for unrecognized visceral burns through capacitively coupled current using monopolar electrosurgery. Plastic cannulas (and reducers) should be avoided with the metal suction-irrigation electrode; an all-metal trocar cannula confers limited safety. With surgeon education and advances in engineering, the potential for unrecognized visceral injury with capacitive coupling can be eliminated. |