Design of a horizontal neutron reflectometer for the European Spallation Source

Autor: J. F. Moulin, Markus Strobl, Klaus Lieutenant, D. Nekrassov, Roland Steitz, M. Trapp
Rok vydání: 2014
Předmět:
Zdroj: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 755:85-96
ISSN: 0168-9002
DOI: 10.1016/j.nima.2014.04.003
Popis: A design study of a horizontal neutron reflectometer adapted to the general baseline of the long pulse European Spallation Source (ESS) is presented. The instrument layout comprises solutions for the neutron guide, high-resolution pulse shaping and beam bending onto a sample surface being so far unique in the field of reflectometry. The length of this instrument is roughly 55 m, enabling $\delta \lambda / \lambda$ resolutions from 0.5% to 10%. The incident beam is focussed in horizontal plane to boost measurements of sample sizes of 1*1 cm{^2} and smaller with potential beam deflection in both downward and upward direction. The range of neutron wavelengths untilized by the instrument is 2 to 7.1 (12.2, ...) {\AA}, if every (second, ...) neutron source ulse is used. Angles of incidence can be set between 0{\deg} and 9{\deg} with a total accessible q-range from 4*10^{-3} {\AA}^{-1} up to 1 {\AA}^{-1}. The instrument operates both in {\theta}/{\theta} (free liquid surfaces) and {\theta}/2{\theta} (solid/liquid, air/solid interfaces) geometry. The experimental setup will in particular enable direct studies on ultrathin films (d ~ 10 {\AA}) and buried monolayers to multilayered structures of up to 3000 {\AA} total thickness. The horizontal reflectometer will further foster investigations of hierarchical systems from nanometer to micrometer length scale, as well as their kinetics and dynamical properties, in particular under load (shear, pressure, external fields). Polarization and polarization analysis as well as the GISANS option are designed as potential modules to be implemented separately in the generic instrument layout. The instrument is highly flexible and offers a variety of different measurement modes. With respect to its mechanical components the instrument is exclusively based on current technology. Risks of failure for the chosen setup are minimum.
Comment: Matched to the version submitted to Nuclear Instruments and Methods A
Databáze: OpenAIRE