Rational use of the PFA-100 device for screening of platelet function disorders and von Willebrand disease

Autor: Suleyman Cuneyt Karakus, Nilgun Sayinalp, Serafettin Kirazli, Semra Dündar, Hakan Goker, Yahya Buyukasik, O.I. Özcebe, Ibrahim C. Haznedaroglu, Düzgün Özatli
Rok vydání: 2002
Předmět:
Zdroj: Blood Coagulation & Fibrinolysis. 13:349-353
ISSN: 0957-5235
DOI: 10.1097/00001721-200206000-00011
Popis: Two hundred and five patients referred for evaluation of platelet functions and 126 healthy controls were tested with the PFA-100 instrument. A cut-off value of 150 s for collagen/epinephrine (CEPI) closure time (CT) produced most acceptable sensitivity (90%), specificity (85.2%), and positive (82.6%) and negative (91.6%) predictivity values for screening of platelet function disorders and von Willebrand disease (vWD). All patients with vWD and Glanzmann thrombasthenia could be detected by PFA-100. Both CEPI and collagen/adenosine diphosphate (CADP) CTs were elevated in all of these cases. Sensitivity of the device was 81.6% for patients with platelet secretion defects. CADP CT was normal in 63.9% of the patients in this subgroup. Specificity (47%) and positive predictivity (57%) of the instrument were diminished in patients with low hemoglobin concentrations. Depending on the results, an algorithm was developed for screening of platelet function disorders and vWD with PFA-100.
Databáze: OpenAIRE