Accuracy of Self-Monitoring of Blood Glucose: Impact on Diabetes Management Decisions During Pregnancy

Autor: M J Henry, S Reinsch, Carol Major
Rok vydání: 2001
Předmět:
Zdroj: The Diabetes Educator. 27:521-529
ISSN: 1554-6063
0145-7217
DOI: 10.1177/014572170102700407
Popis: PURPOSE This study tested the hypothesis that the accuracy of self-monitoring of blood glucose (SMBG) values of patients with diabetes during pregnancy deviates substantially from reference values. METHODS The patients' glucose values were measured on 6 different SMBG meters; reference values were from the HemoCue B Glucose Analyzer. Over a 5-year period, 1973 comparisons between SMBG values and reference values were recorded during clinic visits and used for this study. Data were analyzed for percent of values that varied more than ±10.5% and ±15.5% from the reference value. Out-of-range data at each variance level were analyzed to determine the impact on medical management if decisions were based solely on SMBG values. RESULTS One third of SMBG readings deviated significantly, which could adversely affect treatment for half of these patients if diabetes management was based on SMBG values. At the 10.5% deviation level, 34% of SMBG meter readings were out of range; 54% of these would have implied erroneous treatment. At the 15.5% deviation level, 18% were out of range; 63% of these would have implied erroneous management. CONCLUSIONS The accuracy of home meters should be verified at regular intervals, and SMBG values should not be the sole criterion for diabetes management during pregnancy.