Limiting artifact in CT stereotaxic periventricular procedures

Autor: Patrick Juneau, Robert J. Maciunas
Rok vydání: 1988
Předmět:
Zdroj: Journal of Neurosurgery. 69:459-460
ISSN: 0022-3085
Popis: ✓ One potential source of computerized tomography image artifact during stereotaxic data acquisition is scatter from standard steel-tipped fixation pins as supplied for use with the Brown-Roberts-Wells stereotaxic headframe. In some cases, this can hinder selection of periventricular targets. Replacement of the steel-tipped pins with aluminum-tipped pins, which produce less artifact, helps to facilitate stereotaxic intervention.
Databáze: OpenAIRE