Use of Memory Meters to Measure Reliability of Self Blood Glucose Monitoring
Autor: | S. Till, Peter H. Sönksen, R. Crane, Clara Lowy, I. N. Scobie, C. D. Williams |
---|---|
Rok vydání: | 1988 |
Předmět: |
Adult
Blood Glucose Male medicine.medical_specialty Self blood glucose monitoring Endocrinology Diabetes and Metabolism Endocrinology Internal Medicine medicine Humans Glycosylated haemoglobin Patient compliance Reliability (statistics) Aged Monitoring Physiologic Glycated Hemoglobin business.industry Middle Aged Reflectivity Surgery Self Care Diabetes Mellitus Type 1 Investigation methods Insulin dependent diabetes Emergency medicine Clinical staff Female business |
Zdroj: | Diabetic Medicine. 5:459-462 |
ISSN: | 1464-5491 0742-3071 |
Popis: | Glucose reflectance meters equipped with memories were used in 21 insulin-dependent diabetic patients to assess reliability of self blood glucose monitoring records. Results reported in logbooks showed widespread discrepancies when compared with corresponding meter-held records. Discrepancy rates of at least one in five were observed in 11 patients. Unreliable reporting was usually recognized by clinical staff, and associated with high glycosylated haemoglobin levels and poor technique in measuring blood glucose at home. |
Databáze: | OpenAIRE |