Popis: |
A new technique of vapor generation assisted by a microplasma was proposed for an inductively coupled plasma mass spectrometry (ICP MS). It was found that, by replacing a traditional pneumatic nebulizer with a microplasma (solution anode glow discharge, SAGD), analytical signals of Ag, Bi, Cd, Hg, Pb, Tl, and Zn were improved 8, 4, 13, 13, 9, 10, and 7 times, respectively. The main factor contributing to boosted analytical signal intensities was the higher analyte flux produced by the novel microplasma system. The measurement precision in SAGD-ICP MS was comparable to that achievable for ICP MS (with pneumatic nebulization), and it did not exceed 2%. The detection limits of Ag, Bi, Cd, Hg, Pb, Tl, and Zn in SAGD-ICP MS were 5, 2, 6, 5, 4, 10, and 20 ng L |