Comparison of 90 Y activity measurements in nuclear medicine in Germany
Autor: | Olaf Scheibe, Ole Nähle, Karsten Kossert, Uwe Schwarz, Marion Ehlers, Karen Bokeloh, Klaus Thieme |
---|---|
Rok vydání: | 2016 |
Předmět: |
Quality Control
medicine.medical_specialty 010403 inorganic & nuclear chemistry computer.software_genre Sensitivity and Specificity 01 natural sciences 030218 nuclear medicine & medical imaging 03 medical and health sciences 0302 clinical medicine Reference Values Germany medicine Yttrium Radioisotopes Medical physics Radiometry Syringe Radiation business.industry Reproducibility of Results 0104 chemical sciences Activity measurements Reference values Data mining Nuclear Medicine Laboratories business computer |
Zdroj: | Applied Radiation and Isotopes. 109:247-249 |
ISSN: | 0969-8043 |
DOI: | 10.1016/j.apradiso.2015.11.005 |
Popis: | In 2014, PTB and the company Eckert & Ziegler organized a national comparison exercise to determine the activity of a (90)Y solution. One aim of the comparison was to assess the measurement capability of hospitals and medical practices in Germany. P6-type vials were filled with aliquots of a radioactive (90)Y solution and then sent to 19 participants who were asked to measure the activity in the ampoules as well as in their own standard geometry using syringes. Most of the submitted results have a deviation of less than ±10% from the PTB reference activity when measured in the P6-type vials. The spread is somewhat larger when measured in a syringe geometry. The comparison revealed that some participants have difficulties in applying decay corrections and only a few participants were capable of estimating realistic measurement uncertainties. |
Databáze: | OpenAIRE |